The latest addition to our range of advanced FID analysers, the S4 SOLAR XPLORE is a compact, portable version of our popular S4 SOLAR rack mount analyser. With built-in features such as a detachable tablet with integrated wireless connection, a precision monobloc heated FID and an optional integrated Zero air generator, the S4 SOLAR XPLORE can be fitted with dual FID detectors for simultaneous monitoring of THC, NMHC and methane.
Key Features
• Single or dual detectors
• Integrated zero air generator
• Rugged and portable
• Wireless tablet interface
• Typical warm up 30 minutes
• Precision monobloc heated FID
• Simultaneous monitoring of THC, NMHC & Methane
• Relevant Standards:
- EN12619
- EN15267-3
- EN15267-4
- ISO25140
- Method 25
- Method 25a
- EURO VI/VII
- ARP1256D
- CPA 2015-C174
Technical Specification
MEASUREMENT TECHNIQUE: Flame Ionisation Detector
MEASURING UNITS: PPM or mg user selectable
MEASURING RANGES:
Range A: 0-1000ppm.
User settable to e.g. 0-1ppm, 0-5ppm, 0-10ppm, 0-50ppm,0-100ppm, 0-500ppm, 0-1000ppm.
Resolution:0.01ppm
Range B: 0-10000ppm.
User settable to e.g. 0-10ppm, 0-50ppm, 0-100ppm, 0-500ppm, 0-1000ppm, 0-5000ppm. 0-10,000ppm.
Resolution: 0.1ppm
Range C. 0-100,000 ppm.
User settable, with resolution of 1ppm
RESPONSE TIME:
THC <1.5 secs
CH4 and NMHC <2.5 secs
REPEATABILITY: <1% FSD
OXYGEN EFFECT: <2% of reading from 0% to 21% O2 (H2He)
LINEARITY: +/- 0.5% FSD or 2% of point EN14181 – dc rel : <0.5R2 : >0.99
DRIFT: +/-0.2ppm or 2% range per week, whichever greater
NOISE: +/-0.1ppm or 1% range, whichever greater
TEMPERATURE EFFECT ON ZERO: <0.15% per oC
TEMPERATURE EFFECT ON SPAN: <0.3% per oC
SAMPLE INLET PRESSURE:
With internal sample pump: -0.6 to +0.4bar
Without internal sample pump: +0.2 to +0.5bar
ACCURACY: <0.2% FSD
PRECISION: <1%
DETECTION LIMIT: 0.05mgC/m3
BYPASS FLOW SENSITIVITY: Less than 2% from 1 to 3 L/min
SAMPLE FILTER: Removable 0.4 micron PTFE 7um non removable stainless steel filter for CFID
DISPLAY: Blank or Detachable Screen
SAMPLE CONDITION: 0-200oC (Heated version)0-80oC non-condensing for CFID
FUEL CONSUMPTION:
Single detector:35ml/min H2 or 180ml/min H2He
Dual detectors:70ml/min H2 or 360ml/min H2He
AIR SUPPLY:
Single detector:<1.1L/min
Dual detector:<1.6L/min
OPERATING CONDITIONS: 5-40oC ambient temperature
OUTPUTS:
0-10 Vdc
RS232
Ethernet
TCP/IP
Optional 4-20 mA
POWER REQUIREMENTS:
100 to 250Vac
Optional 24VDC
600W max
REMOTE CONTROL:
AK protocol
Ethernet
Comes with S4i remote software operating system.
SIZE AND WEIGHT: 19” (w) x 133.5 (h) x 530 mm (d)Apx. 22Kg
IP RATING: IP55