The Signal Group 250 Portable cooler dryer removes water vapour from stack and exhaust emission gases, before analysis by ambient analysers.
The 250 features a semiconductor Peltier cooling system with an aluminium cooling block. Fitted into the block is then a high efficiency heat exchanger made of stainless steel. The condensed water is then removed using a continuous peristaltic pump.
The status of the cooler is indicated by an LED system with high temperature alarm, a water alarm and an indicator that the set temperature is being maintained.
Technical Specification
CONFIGURATION: Single channel
INLET: Maximum inlet temperature: 210oC
OUTLET:
Sample:
dewpoint 5oC ± 0.5oC
Hot air:
exhausted to ambient
Condensate:
pumped to drain
FLOW RATE: 3 l/min (maximum)
PLUMBING:
Gas sample(s): 1/4” tube fittings
Condensate drain: 1/4” stainless steel
POWER: Voltage rating: 115 Vac or 230 Vac. Consumption: 180VA maximum
ENVIRONMENT: Ambient temperature: + 5oC to 35oC. Max Humidity: 95% RH non-condensing
FRONT PANEL:
Mains power on/off switch
Each channel has the following LED indicators:
• Status (green)
• High temperature alarm (red)
• Low temperature alarm (red)
• Water alarm (red)
• Power (green)
ALARMS:
High temperature.
Low temperature
Water condensate present
DIMENSIONS: 630mm x 500mm x 400mm
WEIGHT: 15kg