Product Description

The 200SM Series of 19″ rack mounting cooler dryers remove water vapour from Stack and Exhaust  Emission gases, before analysis by ambient analysers. Errors due to water  interference are removed and condensation is prevented from occurring inside  the analysers. The unit is based on a solid state Peltier cooler element  with a continuous peristaltic pump which removes the unwanted water.

The 200SM Series of 19″ rack mounting cooler dryers remove water vapour from Stack and Exhaust  Emission gases, before analysis by ambient analysers. Errors due to water  interference are removed and condensation is prevented from occurring inside  the analysers. The unit is based on a solid state Peltier cooler element  with a continuous peristaltic pump which removes the unwanted water.

Additional information

Applications

, , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , ,

Key Features

• Available in 1, 2 or 3 channel configurations

• Switchable 115/230 Vac.

• Solid state cooler element

• Front panel status indicators

• Compatible with all gas analysers

• Temperature and water alarm for each channel

• High inlet temperature up to 210oC

• High capacity – up to 9 l/min

• Ambient temperature up to 40oC

Documentation

➔ Ancillary Equipment Brochure

➔ Datasheet

➔ User Manual

Technical Specification

INLET:

Maximum inlet temperature: 210oC

OUTLET:

Sample: 

dewpoint 5oC ± 0.5oC 

Hot air: 

exhausted to ambient

Condensate: 

pumped to drain

FLOW RATE:

3 l/min (maximum) per channel

PLUMBING:

Gas sample(s): 

1/4” tube fittings

Condensate drain: 

1/4” stainless steel

POWER:

Voltage rating: 

115 Vac or 230 Vac ± 15% switchable on rear panel

Consumption: 

90 VA + 90 VA per channel (maximum)

ENVIRONMENT:

Ambient temperature range: + 5oC – 40oC

Max Humidity: 95% RH non-condensing

FRONT PANEL:

Mains power on/off switch

Each channel has the following LED indicators:

– Status (green)

-High temperature alarm (red)

-Low temperature alarm (red)

-Water alarm (red)

-Power (green)

ALARMS:

Each channel has three alarms:

-High temperature

-Low temperature

-Water condensate present

Each channel has one alarm output relay. Each alarm output relay has two independent sets of change over contacts.

DIMENSIONS:

19” rach mounting 5U high

Depth behind mounting panel: 533mm

Protrusion in front of: 45mm

Height: 222mm

Width: 435mm

WEIGHT:

10kg + 5kg per channel