3010 MINIFID – Portable heated FID VOC analyser

Product Description

The 3010 MINIFID Portable heated FID VOC analyser is a small, self contained, and fully portable analyser. It is equally convenient for use on the laboratory bench or as a portable on-site measurement tool. This portable FID features the same detection system as the S4 Solar range of FIDs. It is supplied with a 5m heated sample line and built-in controller as standard. The bottle rack may be detached for use with standard, full sized gas cylinders.

Optional Configuration

The 3010 MINIFID Portable heated FID VOC analyser is a small, self contained, and fully portable analyser. It is equally convenient for use on the laboratory bench or as a portable on-site measurement tool. This portable FID features the same detection system as the S4 Solar range of FIDs. It is supplied with a 5m heated sample line and built-in controller as standard. The bottle rack may be detached for use with standard, full sized gas cylinders.

Key Features

• Designed for portability

• Military style design

• Thousands sold globally

• Many still in operation after years of rigorous work

• Easy to use

• Warm-up typically 20 mins

• No microprocessor or software

• Just 5 simple functions

• 0-10v output or optional logger

Relevant Standards

• EN12619 

• EN15267-3 

• EN15267-4

• ISO25140 

• Method 25 

• Method 25a 

• EURO VI/VII 

• ARP1256D 

• CPA 2015-C174 

Documentation

➔ Datasheet

➔ Brochure

➔ User Manual

Technical Specification

SENSITIVITY: 0.1% of range

LINEARITY: 2% of measured value

DRIFT +/-0.2ppm or 2% range per week, whichever greater

NOISE +/-0.1ppm or 1% range, whichever greater

RESPONSE TIME: T90 in 2 seconds

POWER: 110V AC or 220V AC, 500 Watts

HEATED SAMPLE LINE TEMPERATURE: 190 o C

ANALYSER SAMPLING SYSTEM TEMPERATURE 190 o C

WEIGHT WITH GAS BOTTLES 12 Kg (25 lbs)

WEIGHT WITHOUT GAS BOTTLE 9Kg (18 lbs)

DIMENSIONS 210 x 300 x 240mm

SAFETY INTERLOCK Fuel shuts off if flame goes out

SAMPLE INTERLOCK Sample pump and heated line cannot start until analyser is at set temperature

SAMPLE FILTER 0.4 microns (accessible from rear of analyser)

SAMPLE FLOW 0.4-3 l/min. (internal sample pump)

SAMPLE LINE LENGTHS 5 or 10 metres of 6mm Teflon heated to 190 o C controlled by the analyser. Plug in socket is wired so that the 5 and 10 metre lines connect to different contacts. Other lengths up to 15m on request

OPERATING CONTROLS 3 ½ digit LCD display. Temperature reading for heated line & sampling system. Ignition switch. Range select and calibration controls

RECORDER OUTPUTS 0-10 V DC and 4-20 mAmp isolated

DATA LOGGER Built into analyser with software for PC to log to spreadsheet

GAS – QUICK CONNECTS Colour coded for fuel and calibration gas

GAS BOTTLE CAPACITY 1 litre water capacity

FUEL CONSUMPTION H2 35ml/ min

FUEL CONSUMPTION. H2/HE 180ml/min

CAL GAS FLOW RATE 0.4 l/min

ZERO AIR Internally generated with carbon or catalytic scrubber

FLAME AIR Internally generated with carbon or catalytic scrubber

ANALYSER WEATHER PROTECTION IP 20 (IP 55 when fitted with tailored fabric cover)

OPERATING AMBIENT TEMPERATURE +5 to +40 o C